On this page
- Introduction
- Definitions
- Validation of the depuration process
- Transportation
- Pre-depuration storage
- Post-depuration storage
- Depuration water and facility operations
- Laboratories
- Appendix I
- Appendix II
- Appendix III
Introduction
This document provides information for licensed operators that are conducting depuration as part of a decontamination plan under the Management of Contaminated Fisheries Regulations. More specifically, it describes preventive control measures associated with the process of depuration, and is intended to inform and support the development and implementation of a preventive control plan (PCP).
This document covers historical accepted and proven depuration parameters, as well as other example control measures. It is not intended as an exhaustive list of measures. Always ensure that the control(s) chosen are tailored to the uniqueness of your business and shown to be effective for your situation.
Definitions
Please refer to the Canadian Shellfish Sanitation Program manual for definitions.
Validation of the depuration process
- When validating the depuration process, consider the following criteria:
- validate the process with a minimum of 20 lots of shellfish
- use a minimum depuration time of 44 hours
- each lot used for validation has zero hour results greater than or equal to a geometric mean of 230 faecal coliform/100 grams (g), with no sample < 100 fecal coliforms/100 g
- the number and location of samples within the tank to be taken at the beginning, mid, and end point of the depuration process is statistically valid. The intent of the samples is to determine that all locations in the tank facilitate depuration
- the samples are taken over a number of tanks - if the tanks are identical in all characteristics including water flow and tank dimensions
- use 2,300 faecal coliforms/100 g as a maximum zero hour limit for a 44 hour depuration cycle unless the system demonstrates higher zero faecal coliform levels can be depurated successfully during the validation process
- if adding tanks and systems in existing facilities that already have a validated depuration process, consider re-validating (unless it can be demonstrated that all equipment and parameters are identical to the existing system)
- The depuration system is considered satisfactory under defined processing parameters when faecal coliform analyses of samples of depurated shellfish of 20 successive lots meet the targets listed in Table 1A.
- During the validation period, shellfish are only released to the market upon receipt of acceptable final hour bacteriological results as indicated in Table 1B.
Shellfish species | Geometric mean | Upper 10% value Table note 1 |
---|---|---|
Soft clam (Mya arenaria) |
50 | 130 |
Hard clam (Mercenaria mercenaria, Protothaca staminea, Venerupis philippinarum, Nuttallia obscurata) | 20 | 70 |
Blue Mussel (Mytilus edulis) |
20 | 70 |
Oyster (Crassostrea virginica, Crassostrea gigas) | 20 | 70 |
Table note
- Table note 1
-
Upper 10 percent level is where no more than 10 percent of the samples' results used in the evaluation may exceed the value established as the upper 10 percent level for each species.
Transportation
- If the transportation of shellfish harvested from approved shellfish harvest areas and shellfish harvested from restricted areas occurs in the same vehicle or vessel, adequate preventive controls are in place for the identification and the separation of shellfish during transportation.
Pre-depuration storage
Pre-depuration storage in the marine environment
- Prior to depuration, shellfish may be held at the harvest site over the course of a tide cycle, while harvest is occurring. Prior to depuration, shellfish may also be held in near-shore intertidal/subtidal storage areas. Avoid pre-depuration storage during validation in order to avoid altering 0 hour microbiological results from typical harvest areas.
- The pre-depuration storage area is located at a site that is classified as restricted or conditionally restricted, and meets the water quality criteria for depuration as outlined in section 4 of the CSSP Manual.
- Control and oversight is maintained over storage locations, including restricting access. If storage occurs off-site, an adequate security plan is implemented to prevent non-depurated product from entering the food chain.
- Pre-depuration storage area and associated controls are documented in the Preventive Control Plan.
Pre-depuration dry storage
- Pre-depuration dry storage is only undertaken at the licensed depuration establishment.
- The time between harvesting and the onset of depuration is minimized. It should not exceed three days from the date and time of harvest to the date and time of the start of the depuration process.
- In order to avoid thermal shock, storage temperature of shellfish prior to depuration is not:
- greater than the temperature of the process water as documented in a PCP, and/or
- more than 3°C below that documented temperature.
Post-depuration storage
Post-depuration storage in tanks
- For shellfish held in wet storage tanks after depuration and prior to shipping, the water meets the water quality requirements for depuration.
Post-depuration dry storage
- After depuration, shellfish is refrigerated at 4°C or less.
Depuration water and facility operations
Deviations from these guidelines are possible if the depuration process validation studies show that the depuration process consistently yields bacteriologically acceptable product.
Depuration water
- Assess the water sources to ensure that they are safe and suitable for use.
- Verify daily that the water used for depuration has <2 total coliforms/100 mL.
- Ensure the water used meets the following minimum targets prior to depuration:
- a median or geometric mean faecal coliform count not exceeding 88 MPN/100 mL faecal coliforms and not more than 10% of samples exceeding 260 MPN/100 mL faecal coliforms,
- the salinity is ± 20% of the median salinity regime of the area where the shellfish are harvested, unless salinities outside this range are established as a result of the scheduled depuration process evaluation,
- the temperature is adequate to permit normal metabolic activity of shellfish, the limits to be determined by the depuration process validation, and
- there are no undesirable chemicals or other deleterious substances which may affect shellfish filtration rates or the safety of shellfish.
-
The water used is treated when necessary to make it safe for use. Ultraviolet light (UV) systems are the most common marine water treatments used in Canada. Other methods include chlorination/dechlorination and ozonation/deozonation.
For Ultraviolet systems:
- UV lights are monitored for intensity and are replaced as prescribed by the manufacturer,
- filters are used to reduce the turbidity to required levels if, water turbidity exceeds manufacturer's specifications prior to UV treatment. In the absence of manufacturer specifications, turbidity is not greater than 20 Nephelometric Turbidity Units
- turbidity is monitored,
- water flow rate does not exceed the manufacturer's specifications for treatment,
- an automatic shut off is required before or after the ultraviolet system or other means to ensure that untreated water does not enter the tanks in the event of power or ultraviolet system failure, and
- biological filters used in recirculating systems are maintained according to manufacturer's specifications.
- Other criteria to consider for water used in a depuration process:
- for closed or recirculating systems, the ammonia level of process water is below 0.9 ppm,
- the oxygen content is at least 5 ppm or 50% saturation, and
- cease using a water supply that is sourced from an area affected by a biotoxin closure or filter the water supply to remove any toxic phytoplankton using a validated system.
Depuration facility equipment
- Tanks are:
- self-draining to facilitate cleaning,
- able to maintain a minimum flow rate of 107 litres/minute/cubic metre of shellfish. (= 1 U.S. gallon/minute/U.S. bushel),
- constructed so as to provide adequate water flow throughout the tank so that shellfish has adequate access to incoming clean water,
- constructed to ensure they contain water and shellfish at a minimum volume ratio of:
- 4:1 (equivalent to 142 litres of water per 35.24 litres shellfish, or 5 cubic feet of water per U.S. bushel) for soft clam
- water and shellfish at a minimum volume ratio of 6.4:1 (equivalent to 227 litres of water per 35.24 litres shellfish, or 8 cubic feet per U.S. bushel for hard clams (Manila clams, litteneck clams, quahaugs, varnish clams) and oysters
- limits for other species would be determined during validation, and
- constructed such that there is sufficient volume to permit a minimum of 7.6 cm of water clearance around each container.
- Processing containers are of a suitable size and shape to permit:
- a mid-cycle washing of shellfish,
- an adequate flow of water to all shellfish,
- a maximum depth of 10 cm for Manila and littleneck clams, Atlantic oysters, hard shell clams,
- a maximum depth of 20 cm for soft-shelled clams, and
- a maximum depth of 30 cm for Pacific oysters.
The loading criteria for other species are determined by validation.
Shellfish separation
- The handling and wet storage of shellfish harvested from approved areas or conditionally approved areas in the open status occurs only if there is separation from pre-depurated molluscs by harvesting time or shellfish harvest area.
- The procedure to separate shellfish, including washing and handling shellfish, is documented and implemented effectively so that separation of shellfish is maintained.
Handling
- Shellfish are washed and culled to remove dead, broken, or cracked shellfish prior to depuration.
- During depuration, a tank contains only one lot of shellfish.
Mid depuration cycle cleaning
- Tanks and containers of shellfish are thoroughly cleaned at the mid-point of the depuration cycle, in a manner which will not contaminate the shellfish.
Post validation routine monitoring and verification
- As part of PCP implementation, conduct ongoing monitoring and verification of depuration which includes:
- Shellfish samples from every lot at beginning and end point of depuration are tested.
- The minimum Number of samples from each lot at 0 hours and at the end of depuration is determined by:
- the history of the depuration performance,
- the size of the lot depurated,
- the historical variation of contamination of the harvest area or areas, whether spatial or seasonal, and
- the initial levels of contamination.
- If consistently depurating shellfish, analyse five final hour samples when:
- initial faecal coliform counts are greater than or equal to 1,000/100 g,
- receiving shellfish from diverse shellfish harvest areas, or
- receiving shellfish from areas which experience wide fluctuations in contamination over time.
- Five final hour samples are analysed if a modified depuration process of 72 hours is used.
- A depuration facility may reduce the number of zero and/or final hour samples reduced to one if it:
- has high overall depuration performance and harvest areas with consistently low zero hour results, and
- is processing relatively small lots
- If there are higher levels of initial faecal coliforms, deviations in final product results, or incoming product is from a new area, take 5 zero-hour samples.
Number of samples | Shellfish species | Geometric mean not to exceed | One sample may exceed | No sample to exceed |
---|---|---|---|---|
1 | Soft clam | No value | No value | 170 |
1 | Oyster, hard clams, mussel | No value | No value | 100 |
2 | Soft clam | 125 | No value | 170 |
2 | Oyster, hard clams, mussel | 75 | No value | 100 |
3 | Soft clam | 110 | No value | 170 |
3 | Oyster, hard clams, mussel | 45 | No value | 100 |
5 | Soft clam | 50 | 100 | 170 |
5 | Oyster, hard clams, mussel | 20 | 45 | 100 |
10 | Soft clam | 50 | 130 | 170 |
10 | Oyster, hard clams, mussel | 20 | 70 | 100 |
Deviations
- Any lot which shows a final hour faecal coliform count greater than 170/100 g for softshell clams or 100/100 g for all other shellfish is considered as a lot deviation. If two consecutive lots have counts greater than 130/100 g for softshell clams or 70/100 g for all other shellfish, this will indicate a depuration process deviation. In either case, all information pertaining to the deviation is placed in a deviation file.
- For lots that do not meet the final hour limits in Table 1B the following options are available:
- depurate using a validated modified depuration process,
- return to a restricted area meeting the requirements for depuration harvest,
- have shellfish disposed of or used for purposes other than human consumption, and
- return to a restricted area meeting that is approved for depuration harvest. Shellfish are not re-harvested for depuration for at least 14 days.
- In a self-draining system, if a UV shutdown is significant and/or the water level drops below the level of any shellfish in the tank, then the depuration cycle restarts at the beginning of that 22 or 24-hour cycle.
Laboratories
- When microbiological testing of water and shellfish are conducted as part of monitoring, verification and/or validation activities in support of a PCP, the testing is conducted in a laboratory environment with appropriate quality assurance programs. The options include:
- Use of a private lab accredited to the International Organization for Standardization ISO/IEC 17025 standard by either the Canadian Association for Laboratory Accreditation (CALA) or Standards Council of Canada (SCC), where the method for those analyses are included in the scope of accreditation, or
- Testing in an on-site laboratory within the facility , where the laboratory is:
- accredited to the ISO/IEC 17025 standard by either CALA or SCC, or
- assessed by an acceptable independent expert at least every two years.
Independent expert assessment of on-site laboratories
For facilities that engage an independent expert, consider the following.
- Review and keep the evidence and records associated with the assessment, and ensure that it is an appropriate and valid assessment or accreditation of the laboratory.
- Implement any corrective actions from the assessment or accreditation process - within the identified timeline.
- Follow up on any proficiency testing issues.
Independent expert qualifications
- When choosing an independent expert, consider the following:
- their experience through training and/or certification in auditing
- their formal training, education, and background in microbiology (e.g. specialization/ degree/diploma)
- an impartial professional with no real, perceived, or apparent conflict of interest
- Facilities with an on-site laboratory interested in having their laboratory assessed by an independent expert submit:
- the expert's resume, proof of education/ training (e.g. letter, transcript, or certificate)
- a completed conflict of interest questionnaire (Appendix I ) for evaluation to:
- CFIA Science Branch
- National Manager, Laboratory Quality Management and Accreditation
- Email: LAAB@inspection.gc.ca
CFIA will review the information submitted and will advise the facility whether the expert meets the criteria or not.
Assessment
- Ensure the independent expert uses the checklist (Appendix II) to conduct the assessment.
- If the assessment identifies non-conformities:
- take corrective action
- document the corrective actions
- provide evidence of corrective actions to the independent expert within 30 days of the assessment.
- Follow up with the independent expert to confirm whether they determine the corrective actions implemented as acceptable and can advise the facility.
- Provide the original completed checklist and the summary report for the assessment completed by the independent expert (see Appendix III) to the CFIA Science Branch (as per above contact information) as soon as feasible, but no longer than 60 days after the assessment.
- Keep a copy of the assessment report.
Proficiency testing
- Conduct proficiency testing (PT) at least annually for each shellstock and water testing method performed in the laboratory. Potential sources of proficiency testing include:
- USFDA Shellfish PT Program (no cost),
- Canadian Association for Laboratory Accreditation,
- Proficiency Testing | Sigma-Aldrich, and
- AOAC.
- Where participation in proficiency testing is not possible, the laboratory monitors the validity of the test by quantifying a certified reference material and comparing the obtained result to the certified value.
- Satisfactory performance in proficiency testing or results of monitoring using a certified reference material signals that the laboratory continues to operate in compliance with the preventive control plan. If an unsatisfactory result is obtained:
- document the investigation,
- determine the cause,
- implement suitable corrective actions, and
- verify that corrective actions are taken and deemed effective.
- Ensures that the independent expert reviews the laboratory's handling of proficiency testing failures when conducting their assessment.
Appendix I
Conflict of interest questionnaire – independent expert
This questionnaire is intended for use by the licensed depuration facility.
A conflict of interest can be described as any situation in which personal assets, interests or activities affect in any way, or have the potential to appear to affect, the honest, impartial performance of duties or judgement.
- Is the independent expert performing the assessment an impartial professional, capable of giving a true and honest assessment? Yes
No
- Does the independent expert have any potential benefits or personal interests in the outcome of the assessment? Yes
No
- Does the independent expert have any interest in the records or information obtained during the assessment such that a real, perceived or apparent conflict of interest with the assessment may exist or arise? Yes
No
- Has the independent expert agreed to disclose to the facility all personal or professional activities that could place the assessor in a position of a real, perceived or apparent conflict of interest? Yes
No
Signed:
Appendix II
Checklist for the assessment of laboratories in depuration facilities
This checklist is intended for use by the independent expert engaged to evaluate a laboratory conducting testing to monitor the processes in depuration facilities.
Name of Independent Assessor:
Address:
E-mail:
Depuration facility laboratory:
Address:
E-mail:
Date of Laboratory Evaluation:
Laboratory Represented by (list names)
Title (list titles or function)
9222 B Standard Total Coliform Membrane Filter Procedure, Standard Methods for the Examination of Water and Wastewater, 20th ed, American Public Health Association. | |
Multiple Tube Fermentation Technique for Seawater (APHA Recommended Procedures for the Examination of Sea Water and Shellfish, Part III, A) | |
Multiple Tube Fermentation Technique for Seawater using MA-1 (AOAC Official Method 978.23 Fecal Coliforms in Shellfish-Growing Waters) | |
Multiple Tube Fermentation Technique for Shellfish Meats (APHA Recommended Procedures for the Examination of Sea Water and Shellfish, Part III, B) | |
Standard Plate Counts for Shellfish Meats (APHA Recommended Procedures for the Examination of Sea Water and Shellfish, Part III, B) | |
Elevated Temperature Coliform Plate Method for Shellfish Meats | |
MFHPB-19, Enumeration of Coliforms, Faecal coliforms and of E. coli in foods using the MPN method (Health Canada's Health Protection Branch Method, Compendium of Analytical Methods, Volume 2) |
Complete the checklist by indicating Yes (Y), No (N) or Not Applicable (NA) in the appropriate column after observing each listed requirement or the records for the requirement. Provide a description of any observed non-conformances with as much detail as possible (include equipment ID numbers, media lots numbers, sample IDs, dates etc.). If necessary use a separate page to record the information and attach it to the checklist.
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
1.1 | organization of the laboratory | ||
1.2 | staff training requirements/records | ||
1.3 | standard operating procedures/ methods | ||
1.4 | internal quality control procedures for equipment calibration, maintenance, repair, performance checks | ||
1.5 | laboratory safety Table note 2 | ||
1.6 | Included in facility's internal audit or assessment plan. All laboratory activities covered. Records of any findings and corrective actions maintained. | ||
1.7 | Participation in approved PT program annually. Where participation in proficiency testing is not possible, the laboratory is monitoring the validity of the test by quantifying a certified reference material and comparing the obtained result to the certified value. |
Table note
- Table note 2
-
Assessing compliance with specific safety requirements or standards is outside the scope of this assessment.
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
2.1 | Adequate for workload and storage. | ||
2.2 | Clean, well-lighted. | ||
2.3 | Adequate temperature control. | ||
2.4 | Work surfaces are non-porous, easily cleaned and disinfected. | ||
2.5 | air quality < 15 colonies/plate in 15 min |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
3.1 | pH meter has a standard accuracy of 0.1 pH unit | ||
3.2 | pH electrode consisting of pH half-cell and reference half-cell or equivalent (free from Ag/AgClor with an ion exchange barrier preventing the passage of Ag ions into the medium which may affect the accuracy of the pH reading) | ||
3.3 | automatic temperature compensation (ATC) probe or manual temperature adjustment | ||
3.4 | pH meter calibrated daily or as-used and records maintained | ||
3.5 | pH meter calibration requirements- minimum 2 standard buffers, pH 7 and one near the pH of the expected sample or material being measured (pH4, pH 10) Stock standard buffer protected from contamination. | ||
3.6 | Electrode effectiveness determined | ||
3.7 | Balance provides a sensitivity of at least 0.1 g at a load of 150 g | ||
3.8 | Balance verified monthly using NIST Class S or ASTM Class 1 or 2 weights | ||
3.9 | Refrigerator temperatures monitored at least daily and recorded. | ||
3.10 | Refrigerator temperature maintained between 2°C to 4°C | ||
3.11 | Incubator is 35°C +/- 0.5°C. | ||
3.12 | Thermometers are graduated at no greater than 0.5 degree increments. | ||
3.13 | Sufficient number of working thermometers located throughout incubators. | ||
3.14 | Waterbath is at 44.5°C+/- 0.2. | ||
3.15 | Thermometers in waterbaths are graduated in 0.1 degree increments | ||
3.16 | Waterbath has adequate capacity | ||
3.17 | Level in waterbath covers level of liquid in incubating tubes | ||
3.18 | Air incubator/waterbath temperatures are taken daily and recorded | ||
3.19 | Working thermometers tagged with identification, date of calibration, calibrated temperature and correction factor | ||
3.20 | All working thermometers are appropriately immersed | ||
3.21 | A standard thermometer has been calibrated by NIST or one of equivalent accuracy at 0, 35, and 44.5 degrees (45.5 for Elevated Temperature Coliform Plate (ETCP)); calibration records maintained | ||
3.22 | Standard thermometer is checked annually for accuracy by ice point determination; results recorded and maintained | ||
3.23 | Incubator and waterbath working thermometers are checked annually against the standard thermometer at the temperatures at which they're used; records maintained |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
4.1 | Utensils and containers are clean borosilicate glass, stainless steel or other non-corroding material | ||
4.2 | Culture tubes are of a suitable size to accommodate the volume of ingredients and samples | ||
4.3 | Sample containers are made of glass or some inert material (polypropylene) | ||
4.4 | Dilution bottles and tubes are made of borosilicate glass or plastic and closed with rubber stoppers, caps or screw caps | ||
4.5 | Graduations are indelibly marked on dilution bottles or an acceptable alternative is used to ensure appropriate volumes | ||
4.6 | Pipettes are used to inoculate the sample delivering accurate aliquots, have unbroken tips, are appropriately graduated. Pipettes to deliver 1 mL are not larger than 10 mL. Pipettes to deliver 0.1 mL are not larger than 1 mL. | ||
4.7 | Reusable sample containers are capable of being properly washed and sterilized | ||
4.8 | In washing reusable pipettes, a succession of at least 3 fresh water rinses plus a final rinse of distilled/deionized water | ||
4.9 | In washing reusable sample containers, glassware and plasticware, the effectiveness of the rinsing procedure is established annually or when detergent (brand or lot) is changed by the Inhibitory Residue Test as described in the Standard methods for the Examination of Water and Wastewater. Records are kept. | ||
4.10 | Once during each day of washing several pieces of glassware from one batch are tested for residual acid or alkali using a 0.4% aqueous solution of bromothymol blue |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
5.1 | Autoclave(s) are of sufficient size to accommodate workload |
||
5.2 | Routine autoclave maintenance is carried out and records maintained. | ||
5.3 | Autoclave(s) and/or steam generators serviced annually or as needed, and records maintained | ||
5.4 | Autoclave(s) provides a sterilizing temperature as determined weekly using a calibrated working maximum registering thermometer or equivalent | ||
5.5 | An autoclave standard thermometer has been calibrated by NIST or its equivalent at 121°C | ||
5.6 | Working autoclave thermometers are checked against the autoclave standards thermometer at 121°C yearly | ||
5.7 | Spore suspensions used monthly, and results recorded | ||
5.8 | Heat sensitive tape used with each autoclave batch | ||
5.9 | Autoclave sterilization records including length of sterilization, time and pressure | ||
5.10 | For dry heat sterilized materials, the hot air sterilizing oven provides heating and sterilizing temperature in the range 160° to 180°C | ||
5.11 | A thermometer capable of determining temperatures accurately to monitor operation of hot-air sterilizing oven | ||
5.12 | Records of temperatures and exposure times are maintained for hot-air oven | ||
5.13 | Spore strips used quarterly to evaluate effectiveness of sterilization process in hot air oven | ||
5.14 | Reusable sample containers are sterilized for 60 minutes at 170°C in hot air oven or 15 minutes at 121°C in autoclave | ||
5.15 | Sterility of reusable sample containers is determined for each batch/lot | ||
5.16 | Reusable pipettes are stored and sterilized in aluminum or stainless steel canisters; an equivalent method is acceptable. | ||
5.17 | Reusable pipettes (in canisters) are sterilized in a hot-air oven at 170°C for 2 h | ||
5.18 | Sterility of reusable pipettes is determined with each batch/lot. Results recorded. | ||
5.19 | Hardwood applicator transfer sticks are properly sterilized. | ||
5.20 | Spent broth cultures and agar plates are decontaminated by autoclaving for at least 30 min before conventional disposal |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
6.1 | Commercially available media used except for Medium A-1 and modified MacConkey | ||
6.2 | Dehydrated media and components are properly stored in a cool, clean, dry place. | ||
6.3 | Dehydrated media are labelled with date of receipt and date opened. | ||
6.4 | Make-up water is distilled or deionized and exceeds 0.5 megohm resistance or <2 micro-siemens (µS)conductivity at 25°C to be tested and recorded monthly for resistance or conductivity | ||
6.5 | Make-up water is analyzed for residual chlorine monthly and is at a non-detectable level (<0.1 mg/L) Records are maintained. | ||
6.6 | Make-up water is free from trace dissolved metals as determined annually. | ||
6.7 | Make-up water contains <1000 cfu/ml Heterotrophic Plate Count (HPC) as determined monthly | ||
6.8 | Media sterilized according to manufacturer's instructions | ||
6.9 | Volume and concentration of media in tubes are suitable for the amount of sample inoculated | ||
6.10 | Total time of exposure of sugar broths to autoclave temperatures does not exceed 45 min | ||
6.11 | Media sterility, and positive and negative controls are run with each lot of commercially prepared media or run with each batch of media prepared from its components; results recorded | ||
6.12 | Sterile phosphate buffered dilution water or 0.5% peptone water is used as a sample diluent | ||
6.13 | pH of media is determined after sterilization | ||
6.14 | Stored media is labelled with expiration date or sterilization date. | ||
6.15 | All prepared culture media is stored in accordance with manufacturer's recommendations. | ||
6.16 | All culture media is used within the expiration date recommended by the manufacturer. | ||
6.17 | Culture tubes containing precipitate or Durham tubes containing air bubbles are discarded. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
7.1 | All required control organisms available and properly stored and documented | ||
7.2 | Organisms obtained from a suitable source such as ATCC or source able to provide documented characterization Identify Source: |
||
7.3 | Procedure for storing and maintaining control cultures technically valid so suitable controls are always available | ||
7.4 | All working cultures labelled with name and date of subculture |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
8.1 | Containers are of a suitable size to contain at least 100 mL and allow head space for shaking. Seawater samples are collected in clean, sterile, water tight, properly labelled sample containers. | ||
8.2 | Sample identified with collectors name, harvest area, time and date of collection if appropriate | ||
8.3 | After collection, seawater samples shall be immediately placed in a cooler which is maintained between 0-10°C degrees | ||
8.4 | A temperature blank is used to determine the temperature of samples upon receipt at the lab, and results recorded. | ||
8.5 | Examination of the sample is initiated as soon as possible after collection, preferably within 8 hours. However, seawater samples are not to be tested if they are held beyond 30 hours of collection, regardless of refrigeration. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
9.1 | Dehydrated Lawrence Experimental Station (LES) Endo agar or M-Endo medium is used. Manufacturer's directions for rehydration are followed. | ||
9.2 | Sample size of 100 mL is analyzed. | ||
9.3 | Sterile filtration units at the beginning of each filtration series are used. Filtration units are sterilized in accordance with the manufacturer's instructions. | ||
9.4 | Sterile forceps are used to place sterile membrane filter over receptacle. | ||
9.5 | Sample is filter under partial vacuum. Interior surface of the funnel is rinsed with sterile dilution water prior to processing samples. | ||
9.6 | For agar based medium, the membrane filter is placed directly on agar, dish inverted, and incubated for 22-24 hours at 35 +/- 0.5°C. | ||
9.7 | For liquid medium, a pad is placed in the dish and saturated with at least 2 mL of M-Endo and decanted. The filter is placed on the pad, dish inverted and incubated for 22-24 hours at 35 +/- 0.5°C. | ||
9.8 | Positive and negative controls accompany samples throughout the procedure. Records are maintained. Positive Control: Negative Control: |
||
9.9 | For counting, a low power binocular wide field dissecting microscope or other suitable optical device with cool white fluorescent light source is used to provide optimal viewing of sheen. | ||
9.10 | Coliform verification: Monthly verification of at least 10 sheen colonies and a representative number of atypical colonies using the lactose fermentation test or alternative coliform verification tests (using test reactions for cytochrome oxidase and B-galactosidase) is conducted. | ||
9.11 | Results are reported as total coliforms per 100 mL. Where no coliform colonies are observed, report as < 1 coliform/100 mL. For verified counts, the initial count is adjusted based on the positive verification percentage. Results are reported as verified coliform count/100 mL. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
10.1 | Lactose broth or lauryl tryptose broth used as the presumptive media (circle the appropriate one) | ||
10.2 | Sample and dilutions mixed vigorously (25 times in a 30 cm arc in 7 sec) before inoculation | ||
10.3 | In a multiple dilution series 5 tubes per dilution are used | ||
10.4 | For depuration, a single dilution series between 5 and 12 tubes may be used | ||
10.5 | In a single dilution series, the volumes examined are adequate to meet the needs of routine monitoring. Sample volume inoculated: Range of MPN: Strength of media used: |
||
10.6 | Inoculated tubes incubated at 35°C ± 0.5°C for up to 48 ± 3 hours | ||
10.7 | Positive and negative controls accompany samples throughout the procedure. Records are maintained. Positive Control: Negative Control: |
||
10.8 | Inoculated media are read after 24 ± 2 hours and at 48 ± 3 hours and transferred at both intervals if positive for gas. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
11.1 | Brilliant green bile 2% broth (BGB) is used as the confirmatory medium for total coliforms. | ||
11.2 | EC medium is used as the confirmatory medium for fecal coliforms. | ||
11.3 | Transfers are made to BGB/EC either by sterile loop or sterile hardwood applicator stick from presumptive positives incubated for 24 and 48 hours. (Circle the method of transfer) | ||
11.4 | When inoculation is performed using the same loop or applicator stick the order of inoculation is EC first followed by BGB. | ||
11.5 | BGB tubes are incubated at 35 ± 0.5°C. | ||
11.6 | BGB tubes are read after 48 ± 3 hours. | ||
11.7 | EC tubes are incubated in a circulating waterbath at 44.5 ± 0.2°C for 24 ± 2 hours. | ||
11.8 | The presence of any amount of gas or effervescence in a culture tube constitutes a positive test |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
11.9 | Results of multiple dilution tests are read from tables in Recommended Procedures 4th Edition | ||
11.10 | Results from single dilution series are calculated from Hoskins equation or interpolated from figure 1 Public Health report 1621 entitled "Most Probable Numbers for Evaluation of Coli aerogenes Tests by Fermentation tube Method" | ||
11.11 | Results are reported as MPN/100 mL of sample |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
12.1 | Medium A-1 is prepared according to the following recipe (preformulated medium is unacceptable):
Suspend the above ingredients in 1.0 L of distilled or deionized water. Mix thoroughly then add 1 mL of Triton X-100 and continue mixing until dissolved completely. Double strength media is prepared using the above amounts dissolved in 500 mL of water. Dispense 10 mL aliquots into tubes containing inverted Durham tubes. Autoclave at 121°C for 10 minutes. The pH of the medium should be 6.9 after sterilization. |
||
12.2 | Prepared A-1 media is stored in the dark at room temperature and used within 7 days. | ||
12.3 | Sample and dilutions mixed vigorously (25 times in a 30 cm arc in 7 sec) before inoculation | ||
12.4 | In a multiple dilution series 5 tubes per dilution are used | ||
12.5 | For depuration, a single dilution series between 5 and 12 tubes may be used | ||
12.6 | In a single dilution series, the volumes examined are adequate to meet the needs of routine monitoring. Sample volume inoculated: Range of MPN: Strength of media used: |
||
12.7 | Positive and negative controls accompany samples throughout the procedure. Records are maintained. Positive Control: Negative Control: |
||
12.8 | Inoculated media are incubated at 35°C ± 0.5°C for 3 ± 0.5 hours of resuscitation. | ||
12.9 | After 3 ± 0.5 hours resuscitation at 35°C, inoculated media are incubated at 44.5 ± 0.2°C in a circulating water bath for the remainder of the 24 ± 2 hours. | ||
12.10 | The presence of any amount of growth, gas or effervescence in a culture tube constitutes a positive test |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
12.10 | Results of multiple dilution tests are read from tables in Recommended Procedures 4th Edition | ||
12.11 | Results from single dilution series are calculated from Hoskins equation or interpolated from figure 1 Public Health report 1621 entitled "Most Probable Numbers for Evaluation of Coli aerogenes Tests by Fermentation tube Method" | ||
12.13 | Results are reported as MPN/100 mL of sample |
Shellfish samples
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
13.1 | A representative sample of shellstock is collected (min 10-12 live animals) | ||
13.2 | Shellstock is collected in clean, water- proof, puncture resistant containers | ||
13.3 | Shellstock is labelled with collector's name, type of shellstock, the source, the harvest area, date and place (if market sample) of collection | ||
13.4 | Shellstock samples are maintained in dry storage between 0 -10°C degrees until examined | ||
13.5 | Examination of the sample is initiated as soon as possible after collection. However, shellfish samples are not examined if the time interval between collection and examination exceeds 24 hours |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
14.1 | Shucking knives, scrub brushes and blender jars are sterilized (autoclaved) for 15 min prior to use. | ||
14.2 | Blades of shucking knives are not corroded | ||
14.3 | Prior to scrubbing and rinsing debris off shellstock, the analyst's hands are washed with soap and water | ||
14.4 | The faucet used to provide the potable water for rinsing shellstock does not contain an aerator | ||
14.5 | Shellstock are scrubbed with a stiff, sterile brush and rinsed under water of drinking water quality. | ||
14.6 | Prior to opening the molluscs, the hands or gloves are rinsed with 70% alcohol | ||
14.7 | Shellstock are not shucked directly through the hinge | ||
14.8 | Contents of the shellstock (liquor and meat) are shucked into a sterile tared blender jar or other sterile container. | ||
14.9 | At least 100 g of shellfish meat is used for analysis (a minimum of 10-12 live animals) | ||
14.10 | The sample is weighed to the nearest gram and an equal amount by weight of diluent (tempered for ETCP) is added (a 1 in 2 dilution) | ||
14.11 | Sterile phosphate buffered dilution water or 0.5% peptone water is used as the diluent. | ||
14.12 | Sterile phosphate buffered saline is used as diluent for ETCP procedure | ||
14.13 | Samples are blended at high speed for 60 to 120 seconds | ||
14.14 | For other than shellstock APHA Recommended Procedures are followed for examination of freshly shucked or frozen shellfish meats. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
15.1 | Appropriate strength lactose or lauryl tryptose broth is used as presumptive media in the analysis (circle the appropriate broth) | ||
15.2 | Immediately after blending (within 2 minutes) the ground sample is diluted and inoculated into tubes of presumptive media. | ||
15.3 | Must use a 5 tube MPN | ||
15.4 | From the initial 1 in 2 dilution a 1 in 10 dilution is prepared (20 g of 1 in 2 added to 80 g of diluent). From the 1 in 10 a 1 in 100 dilution is prepared (10 g of 1 in 10 added to 90 g diluent). A 5 tube dilution series is inoculated using 10 mL and 1mL from the 1 in 10 dilution and 1 mL from the 1 in 100 dilution | ||
15.5 | In a single dilution series the volumes examined are adequate to meet the needs of routine monitoring Sample volume inoculated: Range of MPN: Strength of media used: |
||
15.6 | Positive and negative control cultures accompany samples throughout the procedure. Records maintained. Positive control: Negative control: |
||
15.7 | Inoculated media are incubated 35°C ± 0.5°C | ||
15.8 | Presumptive tubes are read at 24 ± 2 hours and transferred if positive. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
16.1 | EC medium is used as the confirmatory medium | ||
16.2 | Transfers are made to EC medium by either sterile loop or sterile hardwood applicator sticks. (circle the method of transfer) | ||
16.3 | EC tubes are incubated in a circulating water bath at 44.5°C ± 0.2°C for 24 h ± 2 hours. | ||
16.4 | EC tubes are read for gas production after 24 h ± 2 hours incubation. | ||
16.5 | Presence of any amount of gas or effervescence in Durham tubes constitutes a positive test. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
16.6 | Results of multiple dilution tests are read from tables in Recommended Procedures 4th Ed and multiplied by the appropriate dilution factor | ||
16.7 | Results from single dilution series are calculated from Hoskins equation or interpolated from figure 1-Public Health Report 1621 entitled "Most Probable Numbers for Evaluation of Coli aerogenes Tests by Fermentation tube method" | ||
16.8 | Results are reported as MPN/100 g of sample |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
17.1 | In the standard plate count procedure at least four plates, duplicates of two dilutions are used to provide 30-300 colonies per plate | ||
17.2 | 15 to 20 mL of tempered sterile plate count agar (PCA) is used | ||
17.3 | Agar tempering bath maintains the agar at 44°C to 46°C | ||
17.4 | Not more than 1mL nor less than 0.1 mL of the sample or dilution is plated | ||
17.5 | Samples or sample dilutions to be plated are mixed vigorously (shaken 25 times in a 30 cm arc in 7 sec) before plating | ||
17.6 | Control plates are used to check the sterility of the air, agar and the diluent | ||
17.7 | Solidified plates are incubated at 35°C ± 0.5°C for 48 ± 3 hours inverted and stacked no more than 4 high | ||
17.8 | Quebec Colony Counter or equivalent to provide necessary magnification and visibility for counting plates. | ||
17.9 | A hand tally or equivalent is used for accuracy in counting |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
17.10 | Colony counts are determined in accordance with Part III, A, Sections 4.31 through 4.33 Recommended Procedures 4th Edition | ||
17.11 | Colony Counts reported as ACC/g of sample |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
18.1 | Sample homogenate is cultured within 2 minutes of blending | ||
18.2 | Double strength Modified MacConkey Agar (MMA) is prepared according to the following recipe: Peptone - 34.0 g |
||
18.3 | Hydrated double strength Modified MacConkey Agar is heated to boiling, removed from the heat, and boiled again. This agar is never autoclaved. | ||
18.4 | The twice boiled, double strength Modified MacConkey Agar and sterile phosphate buffered saline are maintained in a tempering bath at 45-50°C until used. Prepared MMA is used the day it is made. | ||
18.5 | The equivalent of 6 grams of the homogenate is placed into a sterile container and the volume is brought up to 60 mL with tempered sterile phosphate buffered saline. | ||
18.6 | 60 mL of the tempered MMA is added | ||
18.7 | Container is gently swirled or rotated to mix the contents which are then distributed evenly over 6 to 8 petri plates. | ||
18.8 | Media and diluent sterility are determined with each use. Results are recorded and records are maintained. | ||
18.9 | To determine media productivity, negative and positive control cultures are pour plated in an appropriate concentration to accompany samples throughout the procedure. Positive control: Negative control: |
||
18.10 | Plates are incubated, inverted stacked no more than 4 high within 3 hours of plating at 45.5°C ± 0.5°C for 18 to 30 h. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
18.11 | Quebec Colony Counter or equivalent is used to provide the necessary magnification and visibility | ||
18.12 | A hand tally or its equivalent is used to aid in counting | ||
18.13 | All brick red colonies greater the 0.5 mm in diameter are totalled over all plates and multiplied by a factor of 16.7 to report results as CFU/100 grams of sample. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
19.1 | For all shellfish, 0.5 % peptone water is used for all dilutions | ||
19.2 | Appropriate strength lauryl tryptose broth is used as presumptive media in the analysis | ||
19.3 | For shellfish, immediately after blending (within 2 minutes) the ground sample is diluted and tubes are inoculated. | ||
19.4 | Must use a 5 tube MPN | ||
19.5 | From the initial 1 in 2 dilution a 1 in 10 dilution is prepared (20 g of 1 in 2 added to 80 g of diluent). From the 1 in 10 a 1 in 100 dilution is prepared (10 g of 1 in 10 added to 90 g diluent). A 5 tube dilution series is inoculated using 10 mL and 1mL from the 1 in 10 dilution and 1 mL from the 1 in 100 dilution. | ||
19.6 | In a single dilution series the volumes examined are adequate to meet the needs of routine monitoring Sample volume inoculated: Range of MPN: Strength of media used: |
||
19.7 | Inoculated media are incubated 35°C ± 0.5°C | ||
19.8 | Presumptive tubes are read at 24 ± 2 hours and transferred if positive. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
19.9 | One loopful from each positive Lauryl Sulphate Tryptose (LST) broth tube is transferred to a tube of Brilliant Green Lactose Bile (BGLB) broth. | ||
19.10 | BGLB broth tubes are incubated at 35°C for 24 +/- 2 h, examined for gas formation and re-incubated for an additional 24 +/- 2 h. | ||
19.11 | Formation of gas during 48 +/- 4 h incubation constitutes a positive confirmed test. | ||
19.12 | MPN of confirmed coliforms per 100 g of shellfish is calculated following the instructions in Appendix D of the Health Canada Compendium of Analytical Methods. |
CSSP requirement | Conforms (Y, N or NA) |
Description of the non-conformance, corrective action required and/or comments | |
---|---|---|---|
19.13 | One loopful of each positive LST tube is transferred to EC broth. | ||
19.14 | EC tubes are incubated in a water bath at 44.5°C for 24 +/- 2 h. | ||
19.15 | MPN of confirmed faecal coliforms per 100 g of shellfish is calculated following the instructions in Appendix D of the Health Canada Compendium of Analytical Methods. |
References:
- American Public Health Association (APHA). 1970. Recommended Procedures for the Examination of Sea Water and Shellfish, 4th Edition. APHA, Washington, D.C.
- American Public Health Association 1984. Compendium of Methods for the Microbiological Examination of Foods, 2nd Edition. APHA, Washington, D.C.
- "Interim Guides for the Depuration of the Northern Quahog, Mercenaria mercenaria." 1968. Northeast Marine Health Sciences Laboratory, North Kingstown, RI.
- Association of Official Analytical Chemists (AOAC). 2000. Official Methods of Analyses of the Association of Official Analytical Chemists. 17th Edition, Chapter 17.305, page 22. AOAC, Arlington, VA.
- U.S. Public Health Service (PHS). 1947. Public Health Report, Reprint #1621. PHS, Washington, D.C.
- American Public Health Association (APHA). 1992. Standard Methods for the Examination of Water and Wastewater, 18th Edition. APHA/AWWA/Water Environment Federation (WEF), Washington, D.C.
- Health Canada's Health Protection Branch, MFHPB-19, Enumeration of Coliforms, Faecal coliforms and of E. coli in foods using the MPN method, Compendium of Analytical Methods, Volume 2.
Appendix III
Summary report template for the assessment of on-site laboratory
This report template is intended for use by the independent expert.
Brief description of the visit -no more than one or two paragraphs
Click here to enter text.
List of non-conformances with outcome/ resolution, date of acceptance, and basis for acceptance (state the evidence or information provided).
Click here to enter text.
Final statement or attestation:
I have assessed this laboratory against the requirements of the checklist in Appendix L: Guidelines for On-Site laboratories at Registered Bivalve Shellfish Establishments. All non-conforming items that were identified and listed in this summary have been corrected.
Signature and date